Publications of Electrical Metrology Department

International and National Journals

L. Callegaro
On strategies for automatic bridge balancing
IEEE Transactions on Instrumentation and Measurement, Vol. 54, no. 2, pp. 529-532, 2005.

Automatic balancing of coaxial ac impedance bridges asks for simple, robust and efficient strategies suitable to be easily implemented on PC controllers. The paper considers the problem of balancing a bridge with n detectors to be equivalent to the problem of finding the root of a complex-valued vector function F. To avoid electrical modeling of the bridge, F is considered generally unknown except at specific points evaluated by experiment. In this framework, efficient root-finding strategies can be employed; in particular, secant method and two generalizations of it to the case n>1 are proposed. Experimental results on a test circuit are given.

G. Marullo-Reedtz, R. Cerri, W. Waldmann, J. Streit, P. Immonen, I. Blanc, F. Raso, T. Funck, B. Schumacher, E. Dierikx, M. Nunes, P. Vrabcek, D. Rudohradský, O. Gunnarsson, K.-E. Rydler, B. Jeanneret, B. Jeckelmann, T. Pulfer, S.S. Tuhran, O. Yilmaz, J. Williams, H. Slinde, K. Lind, J. Nicolas, M. Lindic, E. Flouda, G. Erdos
Comparison EUROMET.EM-K8 of DC voltage ratio: results
IEEE Transactions on Instrumentation and Measurement, Vol. 54, no. 2, pp. 576-579, 2005.

A regional key comparison of DC voltage ratio has been carried out in Europe. The method to evaluate the comparison reference value, the resulting degrees of equivalence for the voltage ratios 1000 V/10 V and 100 V/10 V and the link to the corresponding worldwide key comparison CCEM-K8 are reported.

G. Boella, I. Mihai, G. Marullo Reedtz, P.P. Capra, E. Gasparotto
The IEN CCC bridge to scale the quantized Hall resistance to 1 - Ω standards
IEEE Transactions on Instrumentation and Measurement, Vol. 54, no. 2, pp. 588-591, 2005.

A resistance bridge based on a cryogenic current comparator has been built at IEN. The use of this bridge simplifies the scaling down from the quantised Hall resistance to decadic standard resistors, at the same time reducing the measurement uncertainty. The bridge is described and the result of a comparison at the level of 100 Ω, with the traditional scaling method used at IEN is reported. The two methods agree well within their uncertainties, for which detailed uncertainty budgets are given.

L. Brunetti, E.T. Vremera
New calibration method for coaxial microcalorimeters
IEEE Transactions on Instrumentation and Measurement, Vol. 54, no. 2, pp. 684-687, 2005.

The most critical point in the realization of the electrical power standard based on microcalorimeter is the accurate determination of the microcalorimeter losses and their propagation upon the measurand, i.e. the effective efficiency of the sensor under calibration. At IEN Galileo Ferraris, a procedure has been introduced that makes the microcalorimeter constant determination more accurate and reliable. Experiments with the 7-mm coaxial microcalorimeter in 50 MHz - 18 GHz range have shown a significant accuracy improvement for the Italian national power standard up to 0.2% level.

U. Pogliano, G.C. Bosco
Investigations on the electrometric AC/DC transfer standard
IEEE Transactions on Instrumentation and Measurement, Vol. 54, no. 2, pp. 810-813, 2005.

After about four years of activity, additional investigations have been made on the electrometric ac-dc transfer standard. In particular the ac-dc transfer differences have been checked for the dependence on the separation of the electrodes and for long term drift. Furthermore, the measurements made by means of this instrument have been indirectly compared with the preliminary results of the international comparison CCEM-K9.

L. Callegaro, V. D'Elia, D. Serazio
10 nF capacitance transfer standard
IEEE Transactions on Instrumentation and Measurement, Vol. 54, no. 5, pp. 1869-1872, 2005.

New dielectric capacitance standards with a nominal value of 10 nF, suitable to be employed in the reproduction of the farad unit starting from the quantum Hall resistance, have been constructed. The standards are based on commercial chip capacitance components encased in a thermostat, and are defined as four terminal-pair impedance standards. The capacitance value is within 5-5 relative to nominal; dissipation factor is about 8-5 at 1 kHz. The relative temperature coefficient of the capacitance value versus external ambient temperature is expected to be 1-8 K-1 near 23 ºC.

L. Brunetti, L. Oberto
INRIM (IEN) - Laboratorio di alta frequenza
Tutto_Misure, anno VII, n. 3, p. 225, 2005.

In questo report è segnalata la partecipazione del Laboratorio di Alta Frequenza dell'Istituto Elettrotecnico Nazionale "Galileo Ferraris" a due confronti chiave internazionali CIPM. Si tratta del ciclo CCEM.RF-K19.CL concernente la misura di alti valori di attenuazione tramite attenuatori variabili nella gamma di frequenze 60 MHz - 5 GHz e del ciclo CCEM.RF-K5b.CL riguardante la misura di Parametri di Scattering nella banda 2 - 18 GHz su attenuatori, carichi riflettendi ed adattati in linea coassiale da 7 mm.

Conference Proceedings

L. Brunetti, V. Teppati, D. Parena
Trends of traceability and accuracy assessment in vector network analyzers
Proceedings of the 2 nd International Symposium on Metrology and Physics Application, and Energy Measurement (ISPMPAEM 2004), pp. 181-192, Serpong, Tangeran (Indonesia), August 2004.

Modern Vector Network Analyzers need a complex calibration procedure, aimed to remove the greater part of the errors associated to the hardware and to trace measurements to primary standards. Actually, these primary standards are only indirectly associated to SI fundamental electrical quantities. The system accuracy is mainly related to the uncertainty of the standards, which propagates to the calibration coefficients of the network analyzer, according to the algorithms used by the calibration process. Systematic errors not removed by the calibration process, define a scattering matrix that must be anyway evaluated, for the final measurement uncertainty. Network Analyzer accuracy assessment after calibration process is widely described in literature, but some clarifications are necessary, in order to avoid not consistent interpretations. Indeed, also in high-level comparisons, discrepancies arise among participants using instrumentation of equivalent performances, a result clearly due to different interpretation of the analysis rules. The authors suggest simple principles, in order to improve the harmonization of the measurement results.

L. Callegaro, M. Pisani
A sampling wattmeter for noise power measurements
Proceedings of SPIE International Symposium "Fluctuations and Noise", Vol. 5844, pp. 284-293, Austin (Texas, USA), May 2005 (on cd rom).

Low-frequency noise measurements are usually performed by measuring the voltage across a dipole, or the current through a wire. Here we demonstrate the feasibility of the measurement of the noise power flowing through a line connecting two dipoles. A sampling wattmeter, with spectral display capabilities, having a bandwidth in the range 1 Hz - 10 kHz and sensitivity better than 1-22 W/Hz is here described, and employed to measure the noise power flowing between two resistors at different temperatures. Possible applications of the device include noise thermometry, noise measurement of active devices under load condition, investigations of excess noise below thermal threshold.

P.P. Capra, F. Galliana, F. Francone
Measurement system for the calibration of current clamp-meters: characteristics and traceability levels
Proceedings of the 12 th International Metrology Congress, Lyon (France), June 2005 (on cd rom).

In this paper it is described the measurement system developed at IEN for calibration of clamp-meters in ac current (50 Hz) in the field 0-1500 A. The system is constituted by a source transformer with secondary circuit that in parallel configuration can supply up to 1500 A. The primary coil is controlled by a VARIAC. The supplied current is measured by means of a standard current transformer, a standard resistor of 50 mΩ and a digital voltmeter. The current clamp is applied to a copper wire with suitable section and mounted on a oscillating support. The traceability levels span from 1.5· 10-3 at 100 A to 2.1·10-3 at 50 A to 4.2·10-3 at 1500 A.

E. Vremera, L. Brunetti
Another calibration method for microcalorimeters
14th IMEKO Symposium on New Technologies in Measurement and Instrumentation - 10th Workshop on ADC Modelling and Testing, pp. 71-74, Jurata-Gdynia (Poland), September 2005.

In the paper an original calibration method is described an original calibration for a microcalorimeter based on thermistor power sensors and some preliminary experiments are illustrated, which are necessary for the effective application of the method. This method avoids some critical step in the usual microcalorimeter calibration and does not require vectorial measurements for the feeding path sections.

L. Callegaro, J. Vejdelek
A bridge for high-frequency capacitance scaling
14th IMEKO Symposium on New Technologies in Measurement and Instrumentation - 10th Workshop on ADC Modelling and Testing, oral session, pp. 143-146, Jurata-Gdynia (Poland), September 2005.

A bridge, suitable for 1:1 and 10:1 ratio measurements of four terminal-pair impedances to frequencies up to 5 MHz, has been constructed and is presently under testing. The purpose of the bridge is to extend the Italian national standard of capacitance toward higher frequencies; however, the bridge will be employed for resistance comparisons also. Expected best uncertainty is in the order of a few parts in 10-5 for 10 nF - 1 nF ratio measurements, at 1 MHz.

U. Pogliano, G. La Paglia, D. Serazio
Construction of the traceability for electrical power
14th IMEKO Symposium on New Technologies in Measurement and Instrumentation - 10th Workshop on ADC Modelling and Testing, pp. 321-324, Jurata-Gdynia (Poland), September 2005.

The traceability for the calibration of power converters and wattmeter is obtained in primary laboratory starting from the standards of electrical quantities, such as resistance and voltage. The paper shows the path followed at I.N.Ri.M (which is the primary laboratory in Italy for metrology) to built the traceability for the precision measurement of both for single phase of three-phase active power in the frequency range from 40 Hz to 60 Hz.

E. Vremera, C. Fosalau, L. Brunetti
Instrumentation set-up for characterisation of the sensors based on amorphous wires
14th IMEKO Symposium on New Technologies in Measurement and Instrumentation - 10th Workshop on ADC Modelling and Testing, pp. 458-461, Jurata-Gdynia (Poland), September 2005.

Quality parameters as high sensitivity, quick response, small size, non-contact sense, low-power requirement, make amorphous wires as an attractive material for sensors and microsensors realization. Because other existing non-contact sensors have a good accuracy in many similar applications, the disturbance factors for magneto-impedance sensors based on amorphous wires must be highlighted and cancelled throughout adequate tools. The behaviors of the sensing element must be well characterized through measurements. The paper reviews the most important characteristics for amorphous wires together with the common methods for carry out their most important parameters and proposes an instrumentation set-up based on vector-voltmeter analyzer.

U. Pogliano
A system for the characterization of high resolution DACs and ADCs
14th IMEKO Symposium on New Technologies in Measurement and Instrumentation - 10th Workshop on ADC Modelling and Testing, pp. 563-566, Jurata-Gdynia (Poland), September 2005.

This paper describes the system under construction at I.N.Ri.M. for the characterization of high resolution DACs and ADCs. The main measurement instrument used as a reference is a high precision sampling multimeter, which is properly configured for the measurement of both dc and ac voltages supplied by the DAC under test. A proper voltage pattern is supplied to the ADC under test by means of a multichannel output board put inside a computer, combined with some suitable supplementary electronic circuits. The constructed signal is then measured by the multimeter and analyzed by the computer.

P.P. Capra, F. Galliana
Determination of the pressure coefficients of high precision standard resistors
14th IMEKO Symposium on New Technologies in Measurement and Instrumentation - 10th Workshop on ADC Modelling and Testing (session), Jurata-Gdynia (Poland), September 2005.

In this paper is presented a system, developed at IEN, for determination and measure of pressure coefficients of high precision standard resistors used to maintain the unity of dc electrical resistance, and to transfer the traceability during the periodic international key comparisons established to verify the degree of equivalence of standard among different National metrological Institutes results (NMIs). Preliminary results of the utilisation of the system for the characterisation of high precision standard resistors are also given.

U. Pogliano, G. La Paglia, D. Serazio
Traceability for electrical power at I.N.RI.M.
VI Semetro, invited paper, pp. 20-21, Copacabana, Rio de Janeiro (Brazil), September 2005.

The traceability for the calibration of power converters and wattmeter is obtained at INRiM (Italy) from the standards of electrical quantities (resistance and voltage). The paper shows the path followed to built the traceability for the precision measurement of both for single phase of three-phase active power in the frequency range from 40 Hz to 60 Hz.

L. Brunetti, L. Oberto, E. Vremera
Recent evolutions of the microcalorimeter technique
Proceedings of the XXVIII URSI General Assembly, New Delhi (India) October 2005 (on cd-rom).

The microcalorimeter technique continues to be the key technique for the realization of the primary power standards in the high frequency, because at the state of the art, only it allows tracing the power standards to the direct current standard, that is a SI quantity. This traceability is obtained through the determination of the effective efficiency of a power sensor as a frequency function. Even though since long time the microcalorimeter is a highly explored measurement system, the first realizations being dated from the late of 1950s, still it is possible to propose improvements that should finally increase its accuracy. Actually this one ranges realistically from 0.2% to 2%, if a significant frequency band is considered, e.g. from the radio frequencies to the millimeter waves. Better values are obtainable below 1 GHz, but beyond 18 GHz a 2% still is a challenge, particularly for power sensors in coaxial line of 3.5 mm, 2.92 mm and 2.4 mm. In the paper we highlight the limiting factors of the coaxial microcalorimeter accuracy and how we worked around to improve the technique.

F. Durbiano
Soluzioni acquose di riferimento per la taratura di conduttivimetri
Atti del Congresso Metrologia e Qualità, pp. 69-72, Torino, febbraio 2005.

L'Istituto Elettrotecnico Nazionale Galileo Ferraris (IEN) è da tempo impegnato nell'ambito della metrologia in chimica, in particolare sulla attività riguardante la misura della conducibilità elettrolitica di soluzioni acquose. Il processo di disseminazione è avvenuto sinora attraverso la taratura diretta di misuratori di conducibilità. Utilizzando questa procedura per il trasferimento della riferibilità però si ottiene un'accuratezza della misura limitata dalla periodicità con cui si effettua la taratura, e si priva periodicamente l'utente dello strumento da esaminare. In generale, nell'ambito della metrologia in chimica, una diversa modalità di disseminazione avviene tramite la fornitura di Materiali di Riferimento Certificati (Certified Reference Materials, CRM). I CRM possono essere utilizzati quando non esistono campioni di riferimento o per quelle misurazioni delle proprietà chimiche per le quali il processo è così complesso che l'utilizzo del campione di riferimento porterebbe ad una impossibilità pratica di applicazione. I CRM sono forniti da organismi internazionali specializzati, quali ad es. l'Istituto Europeo per i Materiali e le Misure di Riferimento (IRMM), dagli Istituti Metrologici Primari e da organismi non ufficiali come i produttori di strumentazione. Questi seguono per la definizione, la produzione e la gestione dei CRM le modalità e i requisiti indicati nelle norme e guide riconosciute a livello internazionale. Lo IEN si sta attivando per rendere disponibile il CRM per la taratura dei conduttivimetri. Il CRM è costituito dal flacone contenente la soluzione di riferimento, dal certificato che dichiara il valore di conducibilità e la sua incertezza, e da una procedura con le indicazioni per il corretto utilizzo della soluzione. Questo servizio è offerto per valori di conducibilità compresi tra 1.3 S/m e 50 mS/cm. Questa modalità di effettuare l'operazione di taratura consente sia una migliore accuratezza dovuta alla possibilità di ripeterla in situ a scadenze ravvicinate, sia di avere la strumentazione sempre operativa, sia di eliminare gli effetti del trasporto. Inoltre permette costi diretti del CRM inferiori rispetto a quelli dovuti alla taratura presso lo IEN.

S. Leschiutta
Da quanto tempo non tari il voltmetro?
Atti del Congresso Metrologia e Qualità, pp. 313-315, Torino, febbraio 2005.

U. Pogliano, G.C. Bosco, M. Lanzillotti
Estensione della riferibilità a tensioni alternate inferiori a 300 mV
Atti del Congresso Metrologia e Qualità, pp. 321-323, Torino, febbraio 2005.

C. Cassiago, R. Cerri, G. La Paglia, A. Sosso
Applicazione della linearità di multimetri numerali nel trasferimento di riferibilità in tensione continua
Atti del Congresso Metrologia e Qualità, pp. 324-326, Torino, febbraio 2005.

L' elevata linearità e stabilità dei moderni multimetri digitali (DMM) di alto livello possono essere sfruttate per misurare rapporti di tensione con incertezza molto bassa, effettuando separatamente due misure assolute di tensione. Un multimetro digitale caratterizzato nelle linearità può trovare applicazione anche in laboratori primari, in cui può sostituire il divisore di Kelvin-Varley. Abbiamo sviluppato un metodo integrato per impiegare i multimetri commerciali come campioni di rapporto di tensione per la disseminazione del volt su un ampio intervallo valori: un campione di tensione Josephson costituisce il campione primario di rapporto e uno o più multimetri digitali vengono utilizzati come campioni secondari.

G.C. Bosco, C. Cassiago, R. Cerri, G. La Paglia
Realizzazione di un sistema automatico di taratura in tensione continua basato sul divisore resistivo programmabile IEN
Atti del Congresso Metrologia e Qualità, pp. 327-330, Torino, febbraio 2005.

L'elevata linearità e stabilità dei multimetri numerali (DMM) di alto livello li rende dei validi sostituti dei divisori resistivi Kelvin-Varley nel processo di trasferimento della riferibilità in tensione continua. Sfruttando tale caratteristica lo IEN ha realizzato un sistema di trasferimento completamente automatizzato basato su un DMM HP 3458A, un riferimento allo stato solido Fluke 732B e un divisore resistivo, realizzato allo IEN, costituito da tre serie di resistori a film spesso da 10 kOhm e controllate da un sistema di 10 relè a basse forze termoelettromotrici collegato and un interfaccia GPIB. Il processo di trasferimento con tale sistema prevede una prima fase di caratterizzazione dei rapporti 10:1 e 100:1 del divisore IEN basata sulla linearità del DMM, quindi il trasferimento di riferibilità vero e proprio nell'intervallo 100 mV-1 kV utilizzando il divisore IEN, il DMM e il riferimento a 10 Vcc come apparati campione.

P.P. Capra, F. Galliana, F. Francone
Sistema di misura per la taratura di pinze amperometriche in corrente alternata
Atti del Congresso Metrologia e Qualità, pp. 338-340, Torino, febbraio 2005.

La misura della corrente continua è un'esigenza di molti processi industriali. Impianti galvanici e di elettrodeposizione, di saldatura, gruppi di continuità, generatori di corrente, sistemi utilizzanti serie di batterie richiedono un monitoraggio della corrente per il rispetto di parametri di produzione. Esiste il problema di effettuare misurazioni accurate, rapide e senza modificare gli impianti. A questo scopo vengono utilizzate le pinze amperometriche le quali richiedono periodiche tarature. Si descrive il sistema di misura impiegato all'IEN per misure in corrente alternata nel campo 0 - 1500 A. Il sistema è costituito da un trasformatore di alimentazione nella configurazione in parallelo con cui si eroga fino a 1500 A. La corrente erogata dal primario viene misurata dal complesso TA, resistore campione da 50 mohm e da un multimetro. La pinza amperometrica viene applicata a un conduttore in rame e montata su un supporto basculante. La possibilità di posizionare la pinza in modo variabile consente una migliore caratterizzazione nelle reali condizioni di impiego. Il sistema è automatizzato e consente l'acquisizione dei dati relativi alla corrente applicata in modo automatico.

L. Callegaro, V. D'Elia, J. Vejdelek
Estensione della riferibilità dell'impedenza elettrica a frequenze elevate
Atti del Congresso Metrologia e Qualità, pp. 341-343, Torino, febbraio 2005.

I moderni strumenti commerciali per la misura dell'impedenza spingono la necessità di taratura di campioni di capacità al dominio delle frequenze ultra-acustiche (sino a 10 MHz). L'Istituto Elettrotecnico Nazionale Galileo Ferraris (IEN) lavora alla costruzione di una scala riferita di capacità (1 pF - 1 microF) a queste frequenze. La scala richiede diverse metodologie di taratura, per i campioni di basso valore a dielettrico gassoso e per i campioni di valore più elevato a dielettrico solido.

L. Oberto, L. Brunetti
Analisi delle incertezze negli analizzatori di reti vettoriali
Atti del Congresso Metrologia e Qualità, pp. 402-405, Torino, febbraio 2005.

Gli Analizzatori di Reti Vettoriali, necessitano di una complessa procedura di taratura volta a rimuovere gli errori interni ed assicurare la tracciabilità delle misure a campioni primari. Attualmente i campioni primari più adatti per la taratura sono sezioni di linee di trasmissione caratterizzate da misure dimensionali. L'accuratezza del sistema è principalmente limitata dall'incertezza dei campioni che si propaga, attraverso il processo di taratura dello strumento, in funzione dell'algoritmo usato. La valutazione delle incertezze negli Analizzatori di Reti Vettoriali è ampiamente discussa in letteratura ma la metodologia non è ancora sufficientemente consolidata e pertanto viene messa in evidenza la necessità di sviluppare un metodo comunemente accettato per la valutazione dell'incertezza di misura.

L. Callegaro, J. Vejdelek
Un ponte di rapporto per il confronto di impedenze ad alta frequenza
Atti del GMEE 2005, pp. 9-10, Palermo (Italy), settembre 2005.

E' stato sviluppato un ponte per il confronto di impedenze a frequenze ultra-acustiche. Il ponte permette confronti in rapporto nominale 10:1 di capacità nel dominio 1 pF - 100 nF per frequenze da 100 kHz sino a 5 MHz. Verrà impiegato nell'estensione di frequenza del campione nazionale di capacità.

L. Brunetti, L. Oberto
Avanzamenti nella definizione del campione nazionale di potenza in alta frequenza
Atti del GMEE 2005, pp. 25-26, Palermo (Italy), settembre 2005.

Il continuo lavoro di miglioramento nella definizione del campione nazionale di potenza elettrica in Alta Frequenza (AF) conservato presso l'Istituto Elettrotecnico Nazionale "Galileo Ferraris" (prossimo INRiM) è volto alla diminuzione dell'incertezza ed all'estensione della gamma di frequenze alle quali il campione stesso è disponibile per la disseminazione. In questo lavoro viene presentata la più recente evoluzione del campione che ha prodotto l'estensione in frequenza fino a 40 GHz ed il miglioramento dell'accuratezza.

U. Pogliano, G. La Paglia, D. Serazio
Nuovo laboratorio INRiM per la misura della potenza elettrica
Atti del GMEE 2005, pp. 27-28, Palermo (Italy), settembre 2005.

L. Callegaro, M. Pisani
Un wattmetro per la misura di rumore
Atti del GMEE 2005, pp. 318-324, Palermo (Italy), settembre 2005, relazione a invito.

La misura del rumore elettrico, nel dominio delle frequenze acustiche, è di norma la misura in funzione del tempo o della frequenza della tensione ai capi di un dipolo o, in alternativa, della corrente attraverso un conduttore. Si dimostra qui la possibilità di misurare la potenza di rumore lungo una linea che connette due dipoli, mediante un wattmetro campionatore che è stato appositamente costruito. Termometria, misura di rumore di dispositivi attivi in condizioni di carico, misura del rumore in eccesso al di sotto della soglia del rumore termico sono tra le possibili applicazioni del wattmetro.

Technical Reports

L. Oberto
An algorithm for the determination of the Consensus Value in interlaboratory comparison exercises
IEN Technical Report no. 691, June 2005.

One of the critical points of a measurement comparison among laboratories is the determination of the so called Consensus Value. The main problem is to choose whether the data are consistent or they have to be discarded because of the evidence that the measurand value is different from that expected. In this technical report the problem of the data rejection is analyzed, a possible decision algorithm is presented and it is pointed out its application to the Microwave Power Measurement Comparison AF01.

J. Vejdelek, L. Callegaro
High-frequency impedance bridge
IEN Technical Report no. 692, July 2005.

We report here about the project and development of a new bridge for impedance comparisons in the ultra-acoustic frequency range (from 100 kHz to several MHz). The bridge has a Wheatstone configuration, with a main arm which defines the standards under comparison as four terminal-pair impedances, and a ratio arm composed of gas-dielectric capacitors in two terminal-pair configuration. The bridge is driven by direct digital synthesis generators, and the equilibrium process is under full computer control. At present, the bridge has been tested for the measurement of 100 pF - 1 nF (up to 5 MHz), 1 nF - 10 nF (up to 1 MHz) and 10 nF - 100 nF (now up 100 kHz) capacitance ratios. The ratio arm is composed by 1 pF and 10 pF gas-dielectric capacitors calibrated with a network analyzer technique. In the following, detailed description of the hardware design and software algorithms and procedures and results for the above ratio measurements are presented. The expected relative uncertainty of the bridge is around a few part in 10-5 for frequencies below 1 MHz. A full uncertainty evaluation is under development and will be described in another paper.

F. Durbiano, L. Callegaro, P.P. Capra, V. D'Elia
Key Comparison CCQM-K36: Electrolytic Conductivity at 0.5 and 0.005 S/m, IEN Measurement Report
IEN Technical Report no. 694, October 2005.

In July and August 2005 the first International Key Comparison CCQM-K36 on electrolytic conductivity measurements took place. The purpose of the comparison is to establish the equivalence of measurement of electrolytic conductivity among national metrology institutes and to specify the Key Comparison Reference Value - KCRV for this quantity. In this Technical Report the experimental work carried out at IEN for this Comparison is presented. Results and associated uncertainty budgets are reported for measurements on two different aqueous solutions of HCl, with nominal conductivity of 0.005 S/m, and KCl, with conductivity of 0.5 S/m.

L. Brunetti, L. Oberto, P. Silvestro
Campione di potenza RF nella banda 10MHz - 40GHz: elaborazione automatica dati microcalorimetro
Rapporto Tecnico IEN n. 697, dicembre 2005.

Il campione primario di potenza elettromagnetica è realizzato attraverso un esperimento di microcalorimetria che a partire dalla corrente continua può essere ripetuto in un'ampia banda di frequenza a seconda del tipo di linea di trasmissione con cui si opera. Allo IEN, l'interesse reale è attualmente limitato ad una banda di circa 40 GHz su linea di trasmissione coassiale da 2.92 mm dotata di connettore tipo K. Peraltro si opera comunemente su linea di trasmissione da 7 mm con connettore PCN fino a 18 GHz e su linea da 3.5 mm con connettore PC3.5 fino a 26.5 GHz. L'esperimento, che consiste sostanzialmente nella taratura di un sensore di potenza, richiede la registrazione e la consecutiva analisi di un grande numero di dati. Una corretta opera di filtraggio ed elaborazione in automatico è necessaria per ridurre a tempi accettabili l'operazione di taratura e per ottenere una significativa accuratezza. Nel rapporto sono descritte le operazioni da eseguire sui dati di misura ottenuti durante il processo di realizzazione del campione primario di potenza elettromagnetica nella banda di frequenza 10 MHz - 40 GHz fino ad ottenere il dato finale, cioè il coefficiente di taratura di un sensore di potenza in funzione della frequenza.